top of page

HPLC-Autosampler S5200

X-Y-Z operated sample injecting and sample processing system
The system can be upgraded at any time from the basic version for fixed volume injection with sample loop overfilling, up to variable volume injection mode for variable volumes from 1 μl - 100 μl (optional up to 500 μl) without changing the sample loop.
With the next upgrading step, the S5200 can be used for fully automatic sample processing for precolumn derivatisation. In the derivatisating mode, up to three independent reagents can be added to a specified sample volume, mixed with the sample and programmed with independent reaction times.

In addition, the sample can be diluted with more than 100 times (depending on sample volume and syringe volume). For temperature sensitive samples, the sample tray can be cooled or heated with the optional cooling system.

For chemical inert operation the instrument can be delivered for fully metal-free sample handling (completely equipped with PEEK and FEP materials, including the sample needle).
A wide range of sample racks and sample vials can be used for operation from 1ml up to 5000 ml or by using micro filter plates. Simple operation through large alphanumeric display menu.
   

Injection modes

 

 

1) Fixed volume (standard) with 20 μl sample loop
2) Variable volume (1st grade option) 100 μl sample loop
3) Sample processing (2nd grade option) 100 μl sample loop

Metering device

 

 

 

 

 

 

Standard: 500ml syringe stepper motor driven with a resolution of 35 steps/ 1 μl.
Optional: Syringe resolution
                 100 ml      175 steps / 1 μl
                 250 ml       70 steps / 1 μl
               1000 ml       15 steps / 1 μl
               2500 ml        6 steps / 1 μl
               5000 ml        3 steps / 1 μl

Injection volume

 

Fix         : 1,0 to 9,9 times sample loop volume in 0,1 increments programmable
Variable : 1µl - 5000µl in 1µl increments

Sample processing

 

 

 

3 reagents each one 1 μl to 5000 μl in dependently programmable
3 reaction times independently programmable
needle wash after each processing step programmable
sample mixing after each processing step 1-9 times programmable

Sample diluting

 

up to 5000 μl diluting solvent in 1 μl increments can be added to the sample volume
mixing after diluting 1 - 9 time programmable

Needle wash

 

 

Through purge position or inject position selectable
Wash program for selecting needle wash activity and washing volume 1 - 9 times of the syringe volume

Reproducibility

 

less than 1% upon injection of 10 μl variable volume
less than 0.5% upon 100 μl variable volume and fix volume injection

Memory effect
Injection / Vial
Analysis time

less than 0.01% depending on selected wash program
1 to 9 different volumes programmable for each injection
up to 9999 programmable time base selectable in 1 minute, 0.1 min or in seconds

Sample trays

 

 

 

Standard: 2 pcs. for 60 vials /each total 120 for 1.5ml glass vials or for plastic conical Eppendorf types
Optional: preparative version: 2pcs. for 24 vials /each for 5ml glass vials
micro filter plates holder

Option

 

cooling - heating system minimum: Rt - 20°C up to 70°C
PEEK - Version (chemically inert option including the needle)
Cat.-No.

Description

S5200

Standard Autosampler S5200, fix volume (20µl) injection, max. 120 vials

Cat.-No.

Description

S5200 - 01
S5200 - 02

S5200 - 03
S5200 - 04
S5200 - 013

Variable Injection Volume (1st grade option), 100 μl sample loop, max. 120 vials
Variable Sample Processing (2nd grade option) 60 sample/60 sample processing vials
Cooling (RT- 20°C) / Heating System (max 70°C) upgrade for S5200
Preparative injection upgrade f. S5200, fix volumn 1ml (Standard), max. 48 vials a 5ml
PEEK - Version (chemically inert option, including PEEK - needle)