Operation Principle of the Online-Degasser
In a vacuum chamber, the solvent to be degassed is distributed into several, parallel arranged, semi-permeable membrane chambers.
When a high vacuum (more than -850 mbar) is applied, only the in the liquid dissolved gases are diffusing through the thin membrane.
With a total solvent volume of 8 ml per degassing channel, the Degasser S 7510 achieves an extremely low remaining gas concentration of 0.5 to 5 ppm oxygen, depending on the degasser's flow rate (0 to 10 ml/min.).
HPLC-Online-Degasser
With a given flow rate; the degassing quality can be further increased by coupling two or several membrane channels. The digital display of the actual vacuum in millibar makes the degasser S 7510 the first vacuum degasser on the market which functions can be checked and documented at any time.
The degasser S 7510 is availabel with and without digital display. All functions are the same for both degasser S 7510.
The degasser S 7510 is availabel with and without digital display. All functions are the same for both degasser S 7510.
| Degassing method |
with an applied vacuum, dissolved gases are continuously |
| Max. flow rate | max. 10ml/min. for each degassing channel |
| Max. efficiency | 0,5 ppm oxygen at 0,5 ml/min |
| Parts which come in contact with the solvent | PTFE and PEEK |
| Channels | 2, 3 or 4 channels |
| Volume per each channel | ~ 8ml |
| Power supply | 220/110 V 50/60 Hz |
| Dimensions | 110 x 150 x 350mm |
| Weight | 5,6 kg |
| Cat-no. |
Description |
|
SDS-7510-OD-2 |
2 Channel Degasser without display |
| SDS-7510-OD-3 | 3 Channel Degasser without display |
|
SDS-7510-OD-4
|
4 Channel Degasser without display
|
| SDS-7510-D-2 | 2 Channel Degasser with display |
| SDS-7510-D-3 | 3 Channel Degasser with display |
| SDS-7515-D-4 | 4 Channel Degasser with display |